Machine technical sheet


Immagine rappresentane il macchinario
Brand: RGMD S.p.A. (Genova)
Model: Dibi Skin RF
Type: Beauty treatments - apparatus for the partial treatment of heat through radiofrequency
Power supply: Electrical (220V-380V)
Reference standard: CEI 62-39 - abrogata dal 2016
Frequency of work: 1 - 65000 kHz
Type of emission: Continua
Type of radiant element: Capacitivo
Category according to 12198-1-12009 Standard: Non disponibile
Required protection measures:
  1. Delimit Zone 1: Exposure values above the reference levels for general public
  2. The access to the Restricted areas shall be permitted only to those individuals with proven absence of contraindications to EMF exposure
  3. The access to the Restricted area is forbidden to individuals wearing pacemakers and/or implanted electronic devices

Measurements in the field (Click to view the measurements in the field)

Reference person:

AUSL 7 Siena - Laboratorio Agenti Fisici Nicola Stacchini (n.stacchini@usl7.toscana.it)
SIZE AND SHAPE OF THE APPLICATOR [m]
X 0 SHAPE

Cylindrical
Y -
Z -
Measuring point Reference Distance Perpendicular [cm] Distance Parallel [cm] Height from ground [cm] Frequency dominant [Hz] FIGURE MEASURED ENTITY MEASURED VALUE NOTES
P1
applicatore
10 - - 500000 % reference level General Public council recommendation 1999/519/CE Electrical Stimulation [Electric Field] 95 % Manipolo corpo Potenza 100%
P2
applicatore
10 - - 500000 % reference level General Public council recommendation 1999/519/CE Electrical Stimulation [Magnetic Field] 1 % Manipolo corpo Potenza 100%
P4
applicatore
10 - - 500000 % reference level General Public council recommendation 1999/519/CE Electrical Stimulation [Electric Field] 105 % Manipolo viso Potenza 100%
P8
Torace operatore
10 - - 500000 % reference level General Public council recommendation 1999/519/CE Electrical Stimulation [Electric Field] 95 % Manipolo viso Potenza 100%
P9
Testa Operatore
10 - - 500000 % reference level General Public council recommendation 1999/519/CE Electrical Stimulation [Electric Field] 135 % Manipolo viso Potenza 100%
P9
applicatore
10 - - 500000 Percentage reference level General Public council recommendation 1999/519/CE Thermal Effect [Electric Field] 85 % sul valore massimo Manipolo viso Potenza 100%
P10
applicatore
100 - - 500000 % reference level General Public council recommendation 1999/519/CE Electrical Stimulation [Electric Field] 40 % Manipolo viso Potenza 100%
Zona 0 è la zona all’interno della quale i livelli di esposizione sono sicuramente inferiori o uguali ai livelli di riferimento della popolazione (conformità alla raccomandazione 1999/519/CE) Distanze maggiori di 1 m
Zona 1 è la zona all’interno della quale i livelli di esposizione superano i livelli di riferimento della popolazione ma sono inferiori o uguali ai valori di AZIONE INFERIORI per i lavoratori stabiliti dal Decreto Legislativo n. 159/2016 Distanze minore di 1 m
Zona 2 è la zona nella quale i livelli di esposizione superano i livelli di AZIONE INFERIORI per i lavoratori stabiliti dal Decreto Legislativo n. 159/2016 assente

Note